Home > Peco Food and Beverage > Products> Package Integrity > VeriPac > VeriPac 225 D
VeriPac 225 D
 |
 |
| Application |
Leak detection for packages with defect profile typically >25 microns |
| Package type |
Foil barrier pouches
Poly peelable chevron pouches
Pouch sizes from 1"x3" up to 30"x30" |
| Industries: markets served |
Medical device, pharmaceutical, nutritional, food |
| Package materials and combinations |
All non-porous: Foil, Plastic, Poly, Film, Aluminum |
| On line or off line testing |
Off line lab instrument or On line leak testing applications |
| Test components |
Absolute vacuum transducer |
| Technology |
Vacuum decay |
| Operator interface |
Touch screen |
| Test parameter storage |
Up to 30 packages |
| Test sensitivity |
Manual or automatic |
| Test result data |
Pass/fail result – vacuum decay results measured in Mbar/sec. |
| Statistics and test results trends |
Yes |
| Data collection |
Collects test data for view on touch screen, up to 300 tests.
Electronic data collection (unlimited test data) |
| Test chamber tooling |
Test drawer with large flex membrane for inspection of different size pouches |
| ASTM test method approved |
ASTM F2338* |
| Package defect detected |
Measures whole seal integrity of container or package and quality of entire package |
| Test frame construction |
Stainless steel |
| Instrument dimensions |
20" W – 16" D – 9.5" H (instrument only) 38" W – 24" D – 24" H (with test chamber) |
| Instrument weight |
28 lbs (instrument only) 100 lbs (with test chamber) |
| Power |
100-240 VAC; 50/60 cycles |
| Main air |
Maximum 90 psi |
| Options |
Microcalibrator flowmeter
Validation qualification package (IQ/OQ/PQ) |
| Warranty |
Standard one year parts and labor |
*ASTM Test Method F2338-05 – www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.
Test results may vary according to application and package specifications.
Further information
 |