Home > Peco Food and Beverage > Products> Package Integrity > VeriPac > VeriPac 225 BLV
VeriPac 225 BLV
225 BLV
| Application |
Non destructive leak detection of multi-cavity blister packaging
With defect profile typically >5 microns |
| Package type |
Blister packaging
Semi-rigid, flexible
Package dimensions: maximum size 6"x 8" |
| Industries: markets served |
Pharmaceutical, nutriceutical |
| Test cycle |
< 10 seconds per blister |
| Blister changeover |
Fast changeover from one blister format to another |
| Package materials |
Paper, foil, film, plastic – any combination
Printed or non-printed surface, glossy or flat finish |
| Online or Offline Testing |
Offline lab instrument |
| Test components |
Absolute vacuum transducer
Special camera |
| Technology |
Vacuum decay and special imaging technology (patents pending) |
| Operator interface |
Touch screen Windows based PC |
| Test parameters |
Pre-set parameters for each blister tooling with auto recognition |
| Test result data |
Pass/Fail result – vacuum decay results measured in Mbar/sec. Identifies defective blister cavities in indirect image |
| Statistics and test diagnostics |
Yes |
| Data collection |
PC based data log collection – network ready
Print result data via external printer (option) |
| Test chamber tooling |
Drawer mechanism accommodates tooling inserts for each type/size blister package |
| Package defect detected |
Measures seal and package integrity of entire blister package and identifies defective blister cavities |
| Other notable features |
Based on ASTM vacuum decay leak test method* with addition of vision technology to ID defective blister cavity |
| Test frame construction |
Stainless steel |
| Dimensions |
25" W – 25" D – 36" H |
| Weight |
275 lbs. |
| Power |
100-240 VAC; 50/60 cycles |
| Main air |
Maximum 90 psi |
| Options |
Microcalibrator flowmeter
Validation package (IQ/OQ/PQ) |
| Warranty |
Standard One Year Parts and Labor |
*ASTM Test Method F2338-05 – www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.
Test results may vary according to application and package specifications.
Further information
 |