Home > Peco Food and Beverage > Products> Package Integrity > VeriPac > VeriPac 225
VeriPac 225
 |
 |
| Application |
Leak detection for packages with defect profile typically >25 microns |
| Package type |
Filled and sealed pouches, cups, trays
Empty trays
Lidded flexible and semi-rigid packaging made of Tyvek® or other porous materials |
| Industries: markets served |
Medical device, pharmaceutical, food |
| Package materials and combinations |
Tyvek®, paper, foil, film, aluminum, plastic and laminated materials |
| On line or off line testing |
Off line lab instrument or on line leak testing applications |
| Test components |
Absolute vacuum transducer |
| Technology |
Vacuum decay |
| Operator interface |
Touch screen |
| Test parameter storage |
Up to 30 packages |
| Test sensitivity |
Manual or Automatic |
| Test result |
Data pass/fail result – vacuum decay results measured in Mbar/sec. |
| Statistics and test results trends |
Yes |
| Data collection |
Collects test data for view on touch screen, up to 300 tests.
Electronic data collection (unlimited test data) using Excel
program |
| Test chamber tooling |
Manual or automatic
Can be configured with bladder tooling specifically for Tyvek® applications |
| ASTM test method approved |
ASTM F2338-05* |
| Package defect detected |
Measures whole seal integrity of container or package and quality of entire package |
| Test frame construction |
Stainless steel |
| Dimensions |
20" W – 16" D – 9.5" H |
| Weight |
28 lbs |
| Power |
100-240 VAC; 50/60 cycles |
| Main air |
Maximum 90 psi |
| Options |
Microcalibrator flowmeter
Validation qualification package (IQ/OQ/PQ) |
| High vacuum option |
No |
| Warranty standard |
One year parts and labor |
| Fda listing |
FDA Recognized Consensus Standard #014; Item 168 |
*ASTM Test Method F2338-05 – www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.
Test results may vary according to application and package specifications.